Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
11.8.2
Perform a D1 and D2 density calibration using the Field
Communicator
1.
Read the Prerequistes on
if you have not already done so.
2.
See the following figure.
Measurement support
154