Connection over multidrop network
Figure A-4:
B
A
C
D
A. Signal converter
B.
250–600
Ω
resistance
C. Devices on the network
D. Master device
5.
Start ProLink III.
6.
Choose Connect to Physical Device.
7.
Set Protocol to
HART Bell 202
.
Tip
HART/Bell 202 connections use standard connection parameters. You do not need to
configure them here.
8.
If you are using a USB signal converter, enable Toggle RTS.
9.
Set Address/Tag to the HART polling address configured in the transmitter.
Tips
•
If this is the first time you are connecting to the transmitter, use the default address: 0.
•
If you are not in a HART multidrop environment, the HART polling address is typically left
at the default value.
•
If you are unsure of the transmitter’s address, click Poll. The program will search the
network and return a list of the transmitters that it detects.
10.
Set the PC Port value to the PC COM port that you are using for this connection.
11.
Set Master as appropriate.
Option
Description
Secondary
Use this setting if a primary HART host such as a DCS is on the network.
Using ProLink III with the transmitter
204
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...