Status LED states
(continued)
Table 9-1:
LED state
Alarm condition
Description
Flashing yellow
No alarm
•
Zero calibration procedure is in progress
•
Loop test is in progress
Solid yellow
Low-severity alarm
Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Solid red
High-severity alarm
Alarm condition that will cause measurement
error (outputs in fault)
9.4
View and acknowledge status alerts
The transmitter posts status alerts whenever a process variable exceeds its defined limits
or the transmitter detects a fault condition. You can view active alerts, and you can
acknowledge alerts. Acknowledging alerts is not required.
9.4.1
View and acknowledge alerts using ProLink III
You can view a list containing all alerts that are active, or inactive but unacknowledged.
From this list, you can acknowledge individual alerts or choose to acknowledge all alerts at
once.
1.
View alerts on the ProLink III main screen under Alerts.
All active or unacknowledged alerts are listed, and displayed according to the
following categories:
Category
Description
Failed: Fix Now
A meter failure has occurred and must be addressed immediately.
Maintenance: Fix Soon
A condition has occurred that can be fixed at a later time.
Advisory: Informational A condition has occurred, but requires no maintenance from you.
Notes
•
All fault alerts are displayed in the Failed: Fix Now category.
•
All information alerts are displayed in either the Maintenance: Fix Soon category or the
Advisory: Informational category. The category assignment is hard-coded.
•
The transmitter automatically filters out alerts with Alert Severity set to
Ignore
.
2.
To acknowledge a single alert, check the Ack checkbox for that alert. To
acknowledge all alerts at once, click Ack All.
Transmitter operation
Configuration and Use Manual
127
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...