Prerequisites
The temperature calibration is a two-part procedure: temperature offset calibration and
temperature slope calibration. The two parts must be performed without interruption, in
the order shown. Ensure that you are prepared to complete the process without
interruption. You will need a low-temperature calibration fluid and a high-temperature
calibration fluid. You will not see the effect of the calibration until both the temperature
offset calibration and the temperature slope calibration are complete.
Important
Consult customer support before performing a temperature calibration. Under normal
circumstances, the temperature circuit is stable and should not need an adjustment.
11.10.1
Perform temperature calibration using ProLink III
See the following figure.
Temperature Offset Calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with
low-temperature fluid
Device Tools >
Calibration >
Temperature Calibration >
Temperature Calibration - Offset
Enter temperature of
low-temperature fluid
Done
Temperature Slope Calibration
Start Calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with
high-temperature fluid
Device Tools >
Calibration >
Temperature Calibration >
Temperature Calibration - Slope
Enter temperature of
high-temperature fluid
Start Calibration
Measurement support
Configuration and Use Manual
159
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...