Alert num-
ber
Alert title
Possible cause
Recommended actions
A131
Meter Verification in
Progress: Outputs to
Last Measured Value
A meter verification test is in pro-
gress, with outputs set to
Last Meas-
ured Value
.
•
Allow the procedure to complete.
N/A
Density D1 Calibra-
tion in Progress
A D1 density calibration is in pro-
gress.
•
No action required.
N/A
Density D2 Calibra-
tion in Progress
A D2 density calibration is in pro-
gress.
•
No action required.
N/A
Density D3 Calibra-
tion in Progress
A D3 density calibration is in pro-
gress.
•
No action required.
N/A
Density D4 Calibra-
tion in Progress
A D4 density calibration is in pro-
gress.
•
No action required.
N/A
Zero Calibration in
Progress
A zero calibration is in progress.
•
No action required.
N/A
Reverse Flow
Flow through the device is in the re-
verse direction (against the flow ar-
row).
•
No action required.
12.3
Flow measurement problems
Problem
Possible causes
Recommended actions
Non-zero flow reading
at no-flow conditions
or at zero offset
•
Misaligned piping (especially in new instal-
lations)
•
Open or leaking valve
•
Incorrect sensor zero
•
If the reading is not excessively high, re-
view the live zero. You may need to restore
the factory zero.
•
Check for open or leaking valves or seals.
•
Check for mounting stress on the sensor
(e.g., sensor being used to support piping,
misaligned piping).
•
Contact customer support.
Troubleshooting
170
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...