4.
For density, set Wave Form as desired and enter the required values.
Option
Required values
Fixed
Fixed Value
Sawtooth
Period
Minimum
Maximum
Sine
Period
Minimum
Maximum
5.
For temperature, set Wave Form as desired and enter the required values.
Option
Required values
Fixed
Fixed Value
Sawtooth
Period
Minimum
Maximum
Sine
Period
Minimum
Maximum
6.
Observe the system response to the simulated values and make any appropriate
changes to the transmitter configuration or to the system.
7.
Modify the simulated values and repeat.
8.
When you have finished testing or tuning, disable sensor simulation.
7.1.1
Sensor simulation
Sensor simulation allows you to test the system or tune the loop without having to create
the test conditions in your process. When sensor simulation is enabled, the transmitter
reports the simulated values for mass flow, density, and temperature, and takes all
appropriate actions. For example, the transmitter might apply a cutoff, activate an event,
or post an alert.
When sensor simulation is enabled, the simulated values are stored in the same memory
locations used for process data from the sensor. The simulated values are then used
throughout transmitter functioning. For example, sensor simulation will affect:
•
All mass flow rate, temperature, and density values displayed or reported via
outputs or digital communications
•
The mass total and mass inventory values
•
All volume calculations and data, including reported values, volume totals, and
volume inventories
Complete the configuration
114
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...