12.1
Status LED states
The status LED on the transmitter indicates whether or not alerts are active. If alerts are
active, view the alert list to identify the alerts, then take appropriate action to correct the
alert condition.
Status LED states
Table 12-1:
LED state
Alarm condition
Description
Solid green
No alarm
Normal operation
Flashing yellow
No alarm
•
Zero calibration procedure is in progress
•
Loop test is in progress
Solid yellow
Low-severity alarm
Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Solid red
High-severity alarm
Alarm condition that will cause measurement
error (outputs in fault)
12.2
Status alerts, causes, and recommendations
Alert num-
ber
Alert title
Possible cause
Recommended actions
A001
EEPROM Error
The core processor has experienced
a memory error.
•
Cycle power to the meter.
•
Replace the core processor.
•
Contact customer support.
A002
RAM Error
The core processor has experienced
a memory error.
•
Cycle power to the meter.
•
Replace the core processor.
•
Contact customer support.
A003
No Sensor Response The transmitter is not receiving one
or more basic electrical signals from
the sensor.
This alert often occurs in conjunc-
tion with Alert 102.
•
Check the drive gain and the pickoff
voltage.
•
Check the wiring between the sensor
and the transmitter.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Check the integrity of the sensor tubes.
•
Perform sensor coil resistance checks.
Troubleshooting
162
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...