Table
name
Process fluid
CTL source data
Reference temperature Density unit
53A
Generalized crude and
JP4
Observed density and
observed temperature
15 °C (configurable)
Base density
Range: 610 to
1075 kg/m
3
53B
Generalized products
Observed density and
observed temperature
15 °C (configurable)
Base density
Range: 653 to
1075 kg/m
3
53D
Lubricating oils
Observed density and
observed temperature
15 °C (configurable)
Base density
Range: 825 to
1164 kg/m
3
54C
Liquids with a constant
density base or known
thermal expansion coef-
ficient
User-supplied reference
density (or thermal ex-
pansion coefficient) and
observed temperature
15 °C (configurable)
Base density in kg/m
3
Restriction
These tables are not appropriate for the following process fluids: propane and propane mixes,
butane and butane mixes, butadiene and butadiene mixes, isopentane, LNG, LPG, NGL, ethylene,
propylene, cyclohexane, aeromatics, asphalts, and road tars.
4.8
Set up concentration measurement
This section guides you through loading and setting up a concentration matrix used for
measurement. It does not cover building a concentration matrix.
The concentration measurement application calculates concentration data from process
temperature and density. Micro Motion provides a set of concentration matrices that
provide the reference data for several standard industry applications and process fluids. If
desired, you can build a custom matrix for your process fluid, or purchase a custom matrix
from Micro Motion .
Note
Concentration matrices can be made available on your transmitter either by loading an existing
matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter,
but only one can be used for measurement at any given time. See
Micro Motion Enhanced Density
Application: Theory, Configuration, and Use
for detailed information on building a matrix.
Prerequisites
Before you can configure concentration measurement:
•
The concentration measurement application must be purchased on your
transmitter.
•
The concentration matrix you want to use must be available on your transmitter, or
it must be available as a file on your computer.
Configure process measurement
50
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...