The checksum values for the transmitter firmware and the core processor firmware
must be read during meter commissioning to meet Weights & Measures
requirements for gas applications in Germany. They may also be useful for
MID/Welmec 7.2 test reports.
7.
If required for your installation, monitor the zero value of your system:
a. Choose Device Tools > Configuration > Transmitter Display > Display Variables
and configure Field Verification Zero(FVZ) as a display variable.
b. At the transmitter display, read and record the current value of FVZ.
FVZ is a diagnostic variable that monitors the zero value over a period of 3 minutes.
It must be read during meter commissioning to comply with MID requirements for
Weights & Measures applications in Measuring Instrument Directive (MID)
2004/22/EC. Refer to your Standard Operating Procedures documentation for
detailed instructions.
8.
Enable software security: Device Tools > Configuration > Weights & Measures >
Software Security.
Important
Depending on local requirements, this step may need to be witnessed by a certified Weights
& Measures inspector.
After this step, the transmitter is in a condition of metrological security
(measurement security). The transmitter performs the following actions:
•
Securing all measurement parameters. You can read the current configurations
but you cannot change them.
•
Clearing Status Alert A027: Security Breach.
9.
Install the physical seal.
Important
In most installations, the physical seal is a wire seal that must be installed by a certified
Weights & Measures inspector. The seal is provided by the inspector. The customer is
responsible for supplying a physical setup that will provide an effective location for the wire
seal.
Set up the Weights & Measures application
120
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...