11.2
Use Smart Meter Verification (SMV)
You can run an SMV test, view and interpret the results, and set up automatic execution.
11.2.1
SMV requirements
To use SMV, the transmitter must be paired with an enhanced core processor.
See
for the minimum version of the transmitter, enhanced core processor, and
communication tool needed to support SMV.
Minimum SMV version
Table 11-1:
Item
Minimum version (legacy)
Minimum basic SMV trans-
mitter
Transmitter
6.0
8.0
Enhanced core processor
3.6
4.4
ProLink III
1.0
4.0
Field Communicator
HART device description: de-
vice rev 6, DD rev 2
HART device description: de-
vice rev 8, DD rev 1
If your devices or tools do not meet the minimum version requirements for SMV, you still
may have access to an older version of meter verification if the option was ordered for your
transmitter. See
for a description of the major differences between the older
meter verification and the newer SMV.
Major differences between meter verification and SMV
Table 11-2:
Feature
Meter verification (older)
SMV (newer)
Measurement inter-
ruption
Suspended during the test (3 mi-
nutes)
No interruption necessary
Result storage
No results stored in the transmitter Last 20 results stored in the trans-
mitter (licensed SMV)
Results reporting
Pass/fail/abort
Pass/fail/abort comparison tables
and graphs for stored test results
Test start methods
Manual only
Manual, scheduled, event-based,
Discrete Input
(1) Detailed test analyses such as comparison graphs are not available on the local display.
11.2.2
SMV test preparation
Although it is not necessary to match factory conditions or change the transmitter's
configuration during an SMV test, the test will run more smoothly when conditions are
stable.
Measurement support
142
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...