18
Press <Up>/<Down> arrow to select this
option.
-> Time Calibratn Menu
Press
<Ent>
Calibration
->Esc=Previous Menu
Cal Det to Wash1
Cal Det to Wash2
Cal Det to Wash3
Cal Aux to Wash #1
Cal Aux to Wash#2
Cal Aux to Wash#3
(Note 1)
Press <Up>/<Down> arrow to select this
option.
-> Cal Det to Wash1
(Note 1)
Press <Ent>
(Note 1)
This step is designed to calibrate Formula
1 product to Washer #1: Use a container
that can hold about 24 oz. Press and hold
F7 for at least 2 sec. to start a 5.0 sec.
pump time to Washer #1 (OR press the
Heavy Soil Switch at corresponding
Washer Interface Module). Enter the
actual amount collected in fl. Oz (Note: the
decimal point is pre-programmed, i.e. if
you need to enter 16.4 oz, key in 164 using
the keypad): press <Ent> to change the
cursor to dark blinking square. Key in the
new amount. Press <Ent> again to save.
Press <Esc> when finished.
Note 1: Repeat these steps for Washer #2
and #3 using F8 and F9.
W1 Cal: 5 s.UseF7/Hvy
Ent actual amt:XX.Xoz
(Note 2)
Press <Up>/<Down> arrow to select this
option.
-> Cal Aux to Wash #1
(Note 2)
This step is designed to calibrate Auxiliary
Product to Washer #1: Use a container
that can hold about 24 oz. Press and hold
F7 for at least 2 sec. to start a 20.0 sec.
pump time to Washer #1. OR press the
Heavy Soil Switch at corresponding
Washer Interface Module to calibrate.
Enter the actual amount collected in fl. Oz
(Note: the decimal point is pre-
programmed, i.e. if you need to enter 3.4
oz, key in 34 using the keypad): press
<Ent> to change the cursor to dark blinking
square. Key in the new amount. Press
<Ent> again to save. Press <Esc> when
finished.
Note 2: Repeat these steps for Washer #2
and #3 using F8 and F9.
W1 AUX Cal:20 s.UseF7
Ent actual amt:XX.Xoz