Dräger Medical AG & Co. KGaA
Service Mode Savina Version 3.0
Page 79
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
Setting the calibration classes for the bypass valve (V1) and of the expiratory
valve (V3).
Displaying the internal device temperature.
Displaying the internal, analog measured values of the master processor.
— see
Displaying the external, analog measured values of the master processor.
Displaying the internal, analog measured values of the front processor.
— see
Displaying and calibrating the offset and calibration voltage of the electronics for
the O
2
sensors.
Calibrating the offset voltage of the inspiratory and expiratory pressure sensors.
Displaying the calibration data of the O
2
sensors.
— Up to SW 1.01, not assigned — as of SW 1.10, display of actual status of
display, 7-segment display, and LEDs.
— Up to SW 1.01, not assigned — as of SW 1.10, setting factory default values.
Input of ambient pressure and display of measured pressure values.
In software version 2.0 or higher. Activation of O2 calibration valve.
In software version 2.0 or higher. Display of power supply unit software version.