Error list Savina Version 3.0
Page 121
(UURUOLVW
520IURQWSURFHVVRU)3HUURUJURXS
520IURQWSURFHVVRU)3HUURUJURXS
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\
11.0000
ROM test: Checksum test procedure
faulty.
−
Carry out SW download using current
SW.
−
Replace Control PCB.
11.0100
ROM test: Error found during
checksum start test.
−
Carry out SW download using current
SW.
−
Replace Control PCB.
11.0101
ROM test: Error found during current
checksum test.
−
Carry out SW download using current
SW.
−
Replace Control PCB.
11.0200
Data retention: Read access error.
−
Replace EEPROM FP.
−
Replace RTC.
11.0201
Data retention: Write access error.
−
Replace EEPROM FP.
−
Replace RTC.
11.0210
Data retention: EEPROM read access
error
−