Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
0RQLWRULQJRIDFWXDWRUV
Page 234
1RR[\JHQ
&DXVH
1.
Oxygen supply not connected or damaged.
2.
O2 supply sensor faulty.
'HWHFWLRQ
1.
The input value "O2 supply critical" show too low a supply pressure.
2.
The input value "O2 supply low" show too low a supply pressure.
6XSSUHVVLRQ
1.
The device is in concentrator mode.
5HDFWLRQ
1.
Generation of event "No O2" (warning "!!! O2 supply down").
2.
Message of event "No O2" to routine functions.