Service Mode Savina Version 3.0
Page 99
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
7HVW6WHS
Test step 23 is used to display the offset voltages of the inspiratory pressure sensor (S4.1) and
expiratory pressure sensor (S4.2). This test step is also used to calibrate the offset voltage.
Activate "Calibrate" to start calibration. It is clearly audible when the respective valves (V6.1/V6.2)
switch to ambient air.
Value after calibration: —75 mV to +75 mV
7HVW6WHS
Test step 24 is used to display the following calibration data of the O
2
sensors (S3.1, S3.2):
7HVW6WHS
Up to SW 1.01 -> not assigned.
"Cal.const"
Calibration constant
"Offset voltage"
Electronics offset voltage
"Pressure"
Pressure at which calibration was carried out
"Temperature"
Temperature at which calibration was carried out