Error list Savina Version 3.0
Page 117
(UURUOLVW
0DVWHUSURFHVVRU035$0HUURUJURXS
0DVWHUSURFHVVRU035$0HUURUJURXS
01.1041
Faulty values from data retention:
Calibration of O2 sensor 1.
−
see 01.0308.
01.1050
Faulty values from data retention:
Offset of O2 sensor 2.
−
see 01.0309.
01.1051
Faulty values from data retention:
Calibration of O2 sensor 2.
−
see 01.0310.
01.2000
Data structure in EEPROM older than
newly installed SW.
−
Disappears after new power-on.
01.2001
Data structure in EEPROM newer than
installed SW.
−
Disappears after new power-on.
−
see 01.0300.
−
Carry out SW download using current
SW.
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\
02.0000
Memory test: Error in the internal RAM.
−
Replace Control PCB.
02.0100
Memory test: Error in the external
RAM.
−
Replace Control PCB.
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\