Dräger Medizintechnik GmbH
Repair Procedure Savina Version 2.0
Page 300
5HSDLU3URFHGXUH
5HSODFLQJWKH((3520
• Read out and write down software options (as of SW 1.01) etc.
• Switch off the unit.
• Open the unit (see "
Opening the operator control panel
").
• Remove EEPROM (D40) from the Control PCB.
)LJ
Removing the EEPROM
The information stored in the EEPROM includes operating hours, software options (as of
SW 1.01), device ID (as of SW 1.01), language version, volume, contrast, baud rate,
and measured value display. This information must be read out and written down before
replacing the EEPROM.