Dräger Medical AG & Co. KGaA
Service Mode Savina Version 3.0
Page 88
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
7HVW6WHS
Test step 13 is used to test the nebulizer valve (V5). The valve can be switched on or off. When the
valve is switched on, O2 flows out of the nebulizer socket (8 L/min +5 L/min). The nebulizer is
powered with O2.
7HVW6WHS
"Insp.cal valve" (V6.1)
"Exp.cal valve" V6.2
"Flowinsp"
"Pawinsp"
"Pawexp"
on
off
10 L/min
0 mbar
±0.3 mbar
approx. 40 mbar
off
off
10 L/min
approx. 40 mbar approx. 40 mbar
off
on
10 L/min
approx. 40 mbar 0 mbar
±0.3 mbar
on
on
10 L/min
0 mbar
±0.3 mbar
0 mbar
±0.3 mbar
Connect (short-circuit) the inspiratory socket and the expiratory socket with a ventilation
hose to perform this test step.