18
1.4.11
"Savina Mobil" trolley
1.4.11.1
Castors
Wheels, of which 2 locking. Check screw
fittings for secure seating to 25 Nm.
1.4.11.2
Trolley accessories
Check condition and function of the following
accessories (if fitted):
Humidifier holder (optional)
Cylinder holder (optional)
"Multiple socket outlet" retrofit kit (option)
Battery pack (option, in trolley pedestal)
External battery connecting cable
1.4.12
Unit accessories
Check condition and function of the following
accessories (if fitted):
Hinged arm (if fitted)
Temperature sensor (if fitted)
1.4.13
Compressed gas connecting tubes
Compressed gas connecting hoses comply
with national regulations.
For example, compressed gas connecting
hoses with sealing rings, neutral color
for O
2
NIST connectors, 3 m,
for O
2
NIST connectors, 5 m,
1.4.14
Hose systems
Hose systems correspond to specifications in
the
Instructions for Use/Operating
.
1.4.15
Adult test lung
Check condition and completeness of adult
test lung.
The adult test lung comprises the following
components:
Face mask angle piece
Catheter connector, ISO, size 7
Breathing bag, 2 L
The breathing bag must not be overstretched.
Z F
Z F
Z F
Z
Z D
Z