Error list Savina Version 3.0
Page 149
(UURUOLVW
2FRQFHQWUDWLRQPHDVXUHPHQWHUURUJURXS
2FRQFHQWUDWLRQPHDVXUHPHQWHUURUJURXS
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\
63.0000
O2 sensor no. 2: Sensor not fitted,
faulty, or not operational yet.
−
Replace O2 sensor no. 2.
−
Install O2 sensor no. 2.
−
Allow O2 sensor no. 2 to stabilize.
63.0001
O2 sensor no. 2: Concentration is
outside specification.
−
Check whether the compressed gas at
the O2 connection has a concentration
of 100% vol.
−
Check ambient pressure sensors in
Service Mode, test 27.
−
Calibrate O2 sensor no. 2 to 100 vol.%
O2.
−
Replace O2 sensor no. 2.
63.0004
O2 sensor no. 2: Compensation with
default pressure.
−
Check ambient pressure sensors in
Service Mode, test 27, and adjust if
necessary.