Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
9HQWLODWLRQPRQLWRULQJ
Page 173
6HFRQGSUHVVXUHOLPLW3DZKLJK PEDU
&DXVH
1.
The patient fights the ventilator or is coughing.
2.
The inspiration tube is kinked or obstructed.
3.
One or both pressure sensor(s) deliver(s) too high a measured value.
4.
Pressure source delivers incorrect pressure.
'HWHFWLRQ
1.
The measured value of the inspiratory pressure sensor P
insp
(t) exceeds the value of the
upper airway pressure limit Paw-high + offset for more than 200 ms.
2.
The measured value of the expiratory pressure sensor P
exp
(t) exceeds the value of the upper
airway pressure limit Paw-high
for more than 200 ms.
3.
The measured value of the inspiratory pressure sensor P
insp
(t) exceeds the value of the
upper airway pressure limit Paw-high + offset(t) +5 mbar
for more than 50 ms.
4.
The measured value of the expiratory pressure sensor P
exp
(t) exceeds the value of the upper
airway pressure limit Paw-high +5 mbar
for more than 50 ms.