Dräger Medizintechnik GmbH
Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
9HQWLODWLRQPRQLWRULQJ
Page 192
2.
During standby and 15 s afterwards.
3.
During a detected disconnection and for the period T
apnea
afterwards.
4.
A "Fail to Cycle" alarm has occurred.
5.
An "Airway pressure high" alarm has occurred.
6.
An "Airway pressure low" alarm has occurred.
7.
Hold condition is active (with insp. hold).
8.
Flow monitoring is switched off.
9.
The alarm can be activated by the user under mask ventilation (NIV).
When flow monitoring is switched off, only that part of the monitoring is suppressed which detects
the error by using the expiration start and expiration end information from the expiratory flow
sensor.
5HDFWLRQ
1.
Generation of the event "Apnea" (warning "!!! Apnea")
2.
Transmission of the event "Apnea" to the ventilation control to switch on apnea ventilation, if
applicable.