36
4.11
Testing the contrast
Change the contrast in the "Config" menu.
The contrast of the display changes.
4.12
Testing the micro-filter detection
Remove the micro-filter The micro-filter is
located on the rear panel of Savina behind the
filter cover.
Audible alarm sounds and display message
"!!! No ambient air filter" appears.
Fit micro-filter and cover.
4.13
Testing the Spirolog sensor detection
Remove the Spirolog sensor (flow sensor).
Audible alarm sounds and display message
"!!! Flow Sensor?" appears.
Re-insert the Spirolog sensor.
4.14
Testing the temperature measurement
4.14.1
Test using
Connect the AWT sensor dummy and switch
to "Test" and 34 °C.
Read out temperature value from "Values"
menu of Savina.
Test value: Display = 34 °C
±
1 °C
Remove the AWT sensor dummy.
4.14.2
Testing of temperature sensor (if fitted)
Connect the temperature sensor and take a
reference measurement at room temperature.
Test value:
Permissible deviation between display on
Savina and
: 2 °C
F
F
F
F
F