
All r
ight
s
rese
rved.
Copy
right
reser
v
ed.
R
566
4900F
L1.f
m
03.
12.01
Dräger Medical AG & Co. KGaA
Error list Savina Version 3.0
Page 135
(UURUOLVW
2SHUDWLQJYROWDJHVHUURUJURXS
41.0040
Operating v5 V of the ambient
pressure sensors is too low.
−
Operating v15 V present (see
"Repair Procedure
")?
−
Operating v15 V for O2 supply
pressure sensor on O2 Valve PCB
present (test point T2(+) and T9
(AGND1) on the O2 Valve PCB, see
also chapter "Schematics and
Diagrams
")?
−
Operating v5 V for O2 ambient
pressure sensors on O2 Valve PCB
present (test point T7(+) and T9
(AGND1) on the O2 Valve PCB, see
also chapter "Schematics and
Diagrams
")?
−
Replace the O2 Valve PCB.
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\