Dräger Medical AG & Co. KGaA
Error list Savina Version 3.0
Page 116
(UURUOLVW
0DVWHUSURFHVVRU03UHDGRQO\PHPRU\HUURUJURXS
01.0311
Data retention (EEPROM), checksum
error: Calibration of O2 supply pressure
sensor S5 , ambient pressure sensors
S6 and S7.
−
Carry out Service Mode, test 27
(calibration of ambient pressure
sensors).
01.0312
Data retention (EEPROM), checksum
error: Pneumatics variant (not used for
the time being).
−
01.0313
Data retention (EEPROM), checksum
error (not used for the time being).
−
01.0314
Data retention (EEPROM), checksum
error (not used for the time being).
−
01.1000
Faulty values from data retention: Fan.
−
see 01.0304.
01.1010
Faulty values from data retention:
Blower.
−
see 01.0303.
01.1020
Faulty values from data retention:
Bypass valve calibration class.
−
see 01.0301.
01.1030
Faulty values from data retention:
Expiratory valve calibration class.
−
see 01.0302.
01.1040
Faulty values from data retention:
Offset of O2 sensor 1.
−
see 01.0307.
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\