Alarm Messages Savina Version 2.0
$ODUP0HVVDJHV
9HQWLODWLRQPRQLWRULQJ
Page 184
97LKLJK
&DXVH
1.
Too high a tidal volume is applied due to a change in the state of the lungs.
2.
Spontaneous ventilation of the patient is too high.
'HWHFWLRQ
1.
During a mandatory inspiration (volume-controlled, pressure-controlled) the measured value
VTi(t) exceeds the limit value VTi high set by the user.
2.
During a spontaneous inspiration (pressure-assisted, spontaneous) the measured value
VTi(t) exceeds the limit value VTi high set by the user.
6XSSUHVVLRQ
1.
After a cold start and 15 s afterwards (only alarm).
2.
During standby and 15 s afterwards (only alarm).
3.
During a detected disconnection and 15 s afterwards (only alarm).
4.
The alarm can be activated by the user under mask ventilation (NIV).