Error list Savina Version 3.0
Page 143
(UURUOLVW
*ROG&DSHUURUJURXS
*ROG&DSHUURUJURXS
$X[LOLDU\DODUPHUURUJURXS
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\
51.0000
GoldCap capacity is too low.
−
Replace Control PCB.
51.0001
GoldCap test device is faulty.
−
Replace Control PCB.
51.0002
GoldCap power supply is too high.
−
Replace Control PCB.
−
Carry out SW update
≤
1.10.
51.0003
GoldCap power supply is too low.
−
Replace Control PCB.
51.0004
GoldCap resistance is too high.
−
Replace Control PCB.
(UURUFRGH 'HVFULSWLRQRIPDOIXQFWLRQ
3RVVLEOHUHPHG\
54.0000
Test current for signal generator
(piezo) is too high.
−
Check connection between signal
generator and Control PCB.
−
Replace the signal generator.
−
Replace Control PCB.