Error list Savina Version 3.0
Page 136
(UURUOLVW
2SHUDWLQJYROWDJHVHUURUJURXS
41.0041
Operating v5 V of the ambient
pressure sensors is too high.
−
Operating v15 V present (see
"Repair Procedure
")?
−
Operating v15 V for O2 supply
pressure sensor on O2 Valve PCB
present (test point T2(+) and T9
(AGND1) on the O2 Valve PCB, see
also chapter "Schematics and
Diagrams
")?
−
Operating v5 V for O2 ambient
pressure sensors on O2 Valve PCB
present (test point T7(+) and T9
(AGND1) on the O2 Valve PCB, see
also chapter "Schematics and
Diagrams
")?
−
Replace the O2 Valve PCB.
41.0050
Operating v15 V detected as
incorrect by SiSW.
−
Operating v15 V present (see
"Repair Procedure
")?