Dräger Medical AG & Co. KGaA
Service Mode Savina Version 3.0
Page 94
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
7HVW6WHS
Test step 19 is used to display the measured value of the external A/D converter of the master
processor.
'HVLJQDWLRQ
$'YDOXHV
9ROWDJHUDQJH
5HPDUN
V_FLOW
0016 ... 0049 0.020 V ...
0.060 V)
(with flow exp. = 0). Flow signal of
expiratory flow sensor S2.
V_FLOW_INSP
0246 ... 0491 0.300 V ...
0.600 V)
(with flow insp. = 0). Flow signal of
inspiratory flow sensor S1.
V_PAW_INSP
0491 ... 0737 0.600 V ...
0.900 V)
Measurement signal of the inspiratory
pressure sensor S4.2.
V_PAW_EXP
0491 ... 0737 0.600 V ...
0.900 V)
Measurement signal of the expiratory
pressure sensor S4.1.
V_FIO2_1
0410 ... 0819 0.500 V ...
1.000 V)
(if system is free from O2).
Measurement signal of O2 sensor 1.
The measurement electronics are
located on the O2 Diaphragm PCB.
V_REF_O2
2375 ... 2539 2.900 V ...
3.100 V)
Reference voltage for
. The reference
voltage is generated on the O2
Diaphragm PCB from the battery
voltage.