Dräger Medical AG & Co. KGaA
Service Mode Savina Version 3.0
Page 90
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
7HVW6WHS
Test step 17 is used to display the measured value of the internal A/D converter of the master
processor.
'HVLJQDWLRQ
$'YDOXHV
9ROWDJHUDQJH
5HPDUN
V_FLOW_TEMP
0409 ... 0512 2.00 V ...
2.50 V)
(at ambient temperature). Temperature
of inspiratory gas, measured at TSI
sensor (S1) assembly.
V_TEMP_AIR
0246 ... 0327 1.20 V ...
1.60 V)
(at ambient temperature). Temperature
of inspiratory gas. The temperature
sensor (NTC) is located on the
inspiration block. The measurement
signal is processed on the O2
Diaphragm PCB.
VCC+15V_O2_MESS 0593 ... 0634 2.90 V ...
3.10 V)
Measurement signal of the +15 V
supply voltage of the O2 Diaphragm
PCB/O2 sensors.
VCC+15V_MESS
0573 ... 0655 2.80 V ...
3.20 V)
Measurement signal of the +15 V
supply voltage.