![Diagnostic Instruments SPOT Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/diagnostic-instruments/spot/spot_user-manual_2492839057.webp)
Ch. 4 - Taking a Picture
Captures for Preparation and Modification
User's Guide to the SPOT Cooled Color Digital Camera, 6/9/98
57
To compute new white balance values for an image setup, follow these steps:
1.
Set the illumination and the objective that you will use for the image capture.
2.
Position the slide so that the specimen is not in the field of view. This gives the
camera a reference point to use for a white value. Different specimens require
different points of reference, as described in the following table:
Sample Type
Point of Reference
Transmitted light (non-fluorescent)
Clear part of the slide
Reflected light (non-fluorescent)
Mirror
Macroscopic
White piece of paper
Because fluorescent images are darkfield images (i.e., the background is darker than
the specimen), white balances are not necessary.
3.
From the Camera menu, select the Compute White Balance Values option, or use the
toolbar and click on the Compute White Balance Values button:
4.
The Compute White Balance window appears, as follows:
5.
Click on the Begin button to start the calculation. The Compute White Balance
window displays the status message
Computing white balance values
. You should
hear a series of clicks as the camera samples the light.