Ch. 6 - Calibration and Measurement
Calibration
User's Guide to the SPOT Cooled Color Digital Camera, 6/9/98
105
The Calibration Setup window appears, as shown below:
6.
Enter a name for the Calibration Setup (i.e., 50x).
7.
With the Calibration Setup window open, use the mouse to draw a line along the
length of the stage micrometer image.
a.
Position the cursor on the image (of the stage micrometer).
b.
Left click on the mouse. The drawing function is now enabled. Drag the mouse
across the image of the stage micrometer.
c.
Left click at the point where you want to end the line.
d.
Click on the Calibration Setup screen to re-activate it. The Calibrate button
should now be active.
8.
In the lower boxes, enter the length of the line in whatever units you are calibrating
to, as illustrated below. In this example, the 50x objective has been set to calibrate
the number of pixels for the 5 micron line that you drew: