Ch 3 – Preparing to Take a Picture
Image Setup Options
User’s Guide to the SPOT Cooled Color Digital Camera, 6/9/98
34
When you select the User-Defined option on the Image Setup window (either to create a
new Image Setup or to modify an existing setup, default values appear for the Exposure
Time(s).
To manually enter a new exposure value(s) follow these steps:
1.
Ensure that the User-Defined Exposure box is selected.
2.
Either use the selection arrows or type in a value in seconds for the exposure time.
−
The minimum exposure time is 80 milliseconds (.080 seconds) per color.
−
The maximum exposure time is 17 minutes (1020 seconds) per color.
3.
Click on the OK button to save the settings.
To have the camera calculate the exposure time, follow these steps:
1.
Close the Image Setup window.
2.
Close the Image Setups dialog box.
3.
From the Camera menu, select the Compute Exposure option.
Shortcuts
:
To compute exposure values, you can also use one of the following two
methods:
−
Press
[F 10]
.
−
Click on the Compute Exposure values button on the toolbar:
4.
The Compute Exposure screen appears as shown below: