![Diagnostic Instruments SPOT Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/diagnostic-instruments/spot/spot_user-manual_2492839033.webp)
Ch 3 – Preparing to Take a Picture
Image Setup Options
User’s Guide to the SPOT Cooled Color Digital Camera, 6/9/98
33
6.
When the camera finishes calculating the white balance values, the following window
appears:
7.
The computed white balance values appear, along with the Save to Setup option. The
camera automatically saves the computed values to the current active setup (i.e., the
image setup that appears in the scroll bar at the bottom right corner of the screen).
!
Note: Due to the inherently poor blue light sensitivity of CCD chips and the
energy distribution of halogen bulbs, the blue value is often much higher than the
red and green values.
8.
To change the setup that the new white balance values are saved to, click on Save to
Setup scroll bar, and select a setup from the list.
9.
Click on OK.
10.
The program returns to the main screen.
11.
To edit the white balance values that you saved, open the image setup (i.e.,
Brightfield from the previous example). The values should appear in the Red, Green,
and Blue selection bars. Edit the values as needed.
Exposure Time(s) (User Defined Exposure)
There are two options for entering user-defined exposure times. With the SPOT camera,
you can:
§
Manually enter the value(s).
§
Allow the camera to calculate the value(s).
§
Recall the exposure times and the gain used for the previous image capture.
Each of these options is discussed below.