Appendix B
Changing the Desiccant
User's Guide to the SPOT Cooled Color Digital Camera, 6/9/98
135
8.
Screw the silver plug back into the camera.
Using your fingers, gently start screwing the silver plug back into the camera head,
turning clockwise. Once the threads are started and the Teflon tape is smoothly
turning with the plug, use the 3/8” Allen wrench to finish tightening the plug.
Tighten it well to ensure a good seal.
9.
Place the camera back on the microscope.
Follow these steps:
a)
Remove the plastic lens cap from the front of the camera.
b)
Attach the camera head to the microscope coupler.
c)
Ensure that both your computer and the SPOT power supply are turned off.
d)
Connect the computer cable and the power supply cable to the SPOT camera
head.
!
Note: For a more detailed explanation of installing your SPOT camera, refer to
the box insert that was included with your camera.