3. GENERAL OPERATION
Page 81
© 1998-2008 DH Instruments, a Fluke Company
PG7000
is not able
to establish communication with a valid pressure control component,
<P
control timeout, autogen off>
is displayed momentarily. Correct the communications error
with the pressure control component and try again. If PG7000
is able
to establish
communications with a valid automated pressure control component,
<Turning ON
automated generation>
is displayed momentarily and automated control is turned ON.
The PG7000 automated control function will attempt to float the piston within the piston
position ready limits (see Section 3.4.1) after any
mass to pressure
or
pressure to mass
command once mass loading has been confirmed. It will continue to attempt to float its
piston until
[ENTER]
is pressed to initiate a new command, a function key is pushed to
interrupt AugoGen, or automated pressure control is turned OFF using
[GEN]
.
If automated pressure control is ON and
<1off>
is pressed,
<Turning OFF automated
generation>
is displayed momentarily and automated control is turned OFF. The PG7000
automated control function is inactive.
Automated pressure generation/control ON is indicated by a <G> in the right of the top
line of the PG Terminal main run screen. The <G> flashes when automated generation is active
(see Section 3.7).
Proper operation of the automated pressure generation/control function is dependent on
PG7000’s measurements of piston position. When using AutoGen, be sure that the piston
position indication system is properly adjusted (see Section 5.2.2).
Proper operation of the automated pressure generation/control function is dependent on
PG7000’s measurements of piston position and rotation rate. These measurements are only
available when the mass loading bell is installed on the piston. Automated pressure
generation/control is automatically turned off when the PG7000 entered pressure or mass load
does not include the mass loading bell.
Automated pressure generation/control is not available when operating in PG7000 high line
differential pressure mode (see Section 3.9.4.2).
To help protect against accidental overpressure, when using automated pressure
generation/control, set the upper limit (UL) of the pressure control system using [GEN], <4UL>
(see Section 3.9.9.3).
3.9.9.1
<2TARGET>
PURPOSE
To adjust the distance from midstroke to which the piston position must be set
before the automated generation system considers the piston floating process
complete (see Sections 3.9.9, 3.5).
With PG7202, AutoGen pressure control using the PPCH-G pressure control
does not stop when the target piston position is reached. The PPCH-G thermal
pressure control unit (TPCU) is used continuously to attempt to maintain the piston
at the target position. The default target piston position for a PG7202 is mid-float
position (0.0 mm).
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...