PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 4
•
Storage and one step activation of metrological data on up to 18 piston-cylinder
modules, (3) mass sets and (3) mass loading bells.
•
Continuous
pressure
Ready/Not
Ready indication based on measured
conditions.
•
Motorized, intelligent piston drive system based measured rotation rate with
operator alert and manual override (motorization optional on PG7102 and
PG7302).
•
Integrated automated pressure control with standard
DHI
pressure
controllers.
•
Full RS232 and IEEE-488 communications with multi-level commands to set
and read all instrument functions.
1.2.1.2
AMBIENT AND INSTRUMENT CONDITION
MEASUREMENTS
Temperature
Range
Resolution
Measurement Uncertainty
Ambient
Piston Cylinder Module
0 to 40
o
C
0 to 40
o
C
0.1
0.01
±
1
±
0.1
Barometric Pressure
with Internal Sensor
Range
Resolution
Measurement Uncertainty
70 to 110 kPa
10 Pa
±
140 Pa
Barometric pressure can also be read automatically with any
RS232 device such as a
DHI
RPM.
Relative Humidity
Range
Resolution
Measurement Uncertainty
5 to 95 % RH
1 % RH
±
10 % RH
Piston Position
Range
Resolution
Measurement Uncertainty
±
4.5 mm
0.1 mm
±
0.2 mm
Piston Rotation
(Rate and deceleration)
Range
Resolution
2 to 99 rpm
1 rpm
Vacuum
(PG7601 only)
Range
Resolution
Measurement Uncertainty
0 to 20 Pa
0.01 Pa
±
0.1 Pa or 10 % of reading, whichever is greater
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...