PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 110
Select
<2yes>
to save the all changes
made to the mass loading bell and exit.
Select
<2no>
to abandon all changes
made to the mass loading bell and exit.
Save bell S/N 2832?
1no 2yes
3.11.1.12
EDIT MASS LOADING BELL
PURPOSE
To edit information contained in an existing mass loading bell file. Use this
function to change mass loading bell data after a recalibration.
PRINCIPLE/OPERATION
Mass loading bell information is element specific metrological data.
Uninformed or accidental altering of this information may lead to out of tolerance
measurements. Mass loading bell information should only be edited by qualified
personnel as part of the PG7000 calibration or recalibration process.
See Section 3.11.1.11 describing adding a new mass loading bell. The principles
and procedures to add or edit a mass loading bell are identical except that the mass
loading bell to edit is selected from the existing mass loading bell screen.
3.11.1.13
VIEW MASS LOADING BELL
PURPOSE
To view information contained in an existing mass loading bell file.
OPERATION
To view information contained in an existing mass loading bell file press
[SPECIAL]
and select
<1PC/MS>
,
<3mass bell>
,
<1view>
. The viewing
function is identical to the add function (see Section 3.11.1.11).
3.11.1.14
DELETE MASS LOADING BELL
PURPOSE
To delete an existing mass loading bell file entirely.
OPERATION
Once a mass loading bell file has been deleted it cannot be recovered.
To recreate it, the add mass loading bell function must be used (see Section 3.11.1.11)
which requires entering all the mass loading bell information. Before deleting a mass
loading bell, consider editing it (see Section 3.11.1.12).
To delete a mass loading bell press
[SPECIAL]
,
<1PC/MS>
and select
<3mass
bell>
,
<4delete>.
Confirmation to delete is requested. Select
<2yes>
to delete the mass loading bell.
Press
<1no>
to NOT delete.
The active mass bell (mass bell that is currently selected) cannot be deleted.
Before attempting to delete a mass bell, be sure it is not currently selected.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...