To operate in differential mode and/or execute the RPM offset determination
procedure, the AtmP selection under SETUP must be external (COM2) (see Section
3.10). If this is not the case, an error message will be displayed (<Dif mode requires
COM2 AtmP SETUP>) when attempting to activate differential mode.
To select differential mode operation and/or execute the RPM offset determination
procedure, press
[MODE]
and select
<4dif>
,
<1run>
. The display is:
1.
Current value of RPM offset. Always in Pascal
[Pa].
Offset: 9.03 Pa
New offset? 1no 2yes
Select
<1no>
to use the existing RPM offset and return the previous run screen
in differential mode (see Section 3.9.4.1, Operating in Differential Mode).
The RPM offset should be redetermined any time the static pressure value is
changed, as well as at the beginning of each calibration or test sequence. If the
difference in the value of the RPM offset in subsequent determinations is < 0.1 Pa,
then less frequent offset determinations may be considered.
Select
<2yes>
to proceed with a new determination of the RPM offset
(see Section 3.9.4.1,
PRINCIPAL
. The display is:
1.
Current pressure reading of the RPM in the
current units of.
Set RPM P & [ENT]
96.57785 kPa a
Table 16.
Valve Settings for Setting Differential Mode Static Pressure
VALVE
VALVE SETTING
Control Selection
Static P
PG Shutoff
Closed
Test Bypass
Open
Static P Selection
Off
If static pressure is atmospheric
pressure VENT or OFF
With the Differential Mode Controller valves (see Figure 9) set as indicated
in Table 16 use the system’s pressure control component to set the desired static
pressure value as read by the RPM.
If the desired static pressure is atmospheric pressure (for example, for low gauge
and negative gauge differential pressures), set the STATIC P SELECTION valve
to VENT.
If a REFERENCE VOLUME is connected to the STATIC P port and the desired
static pressure is less than atmospheric pressure, consider setting the STATIC P
SELECTION VALVE to SUPPLY. This will connect the pressure setting vacuum
source directly to the reference volume and reduce pull down time.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...