6. TROUBLESHOOTING
Page 203
© 1998-2008 DH Instruments, a Fluke Company
SYMPTOM
PROBABLE CAUSE
SOLUTION
Pressure defined by PG7000 is
incorrect by relatively small amounts.
A pressure head is applied incorrectly or
inadvertently, the value of gravity used by
the SETUP file is incorrect, the PG7000
Platform is not level, information in the
piston-cylinder module and/or mass set file
is incorrect.
Check and correct if necessary. 3.9.7, 3.10,
2.4.11, 3.11.1
Pressure defined by PG7000 is
incorrect.
Piston is in spring zone although piston
position indicates it is floating freely.
Adjust piston position measurement system.
5.2.2
Cannot set a hydraulic pressure >110
MPa (16 000 psi) with PG7202.
PG7202 believes a gas piston-cylinder
(PC-7200-x) is in operation and gas
piston-cylinder modules cannot be used
above 110 MPa (16 000 psi).
Install and select an oil (PC-7300-x) piston-
cylinder.
When using automated pressure
control pressure generation stopped
and the controller is beeping.
Pressure reached controller’s UPPER
LIMIT.
Check [GEN] UPPER LIMIT and adjust if
desired. 3.9.9.3
When using AMH automated mass
handler, PG7000 still prompts for
manual mass loads.
Mass loading resolution is set to finer than
0.1 kg and prompts are to load trim mass,
AMH mass handler did not initialize.
Adjust mass loading resolution to 0.1 kg,
initialize AMH mass set by selecting it.
3.9.10, 3.11.1.10
The pressure controller TOLERANCE
function setting is too small or the volume
setting (PG7302/PG7307 only) is incorrect.
Adjust the tolerance setting, adjust the
volume setting (PG7302/PG7307 only).
3.9.9.4, 3.9.9.6
Piston position detection system is not
properly calibrated and piston is not
floating as expected.
Calibrate piston position indication system.
5.2.2
Automated pressure control (GEN) is
overshooting the pressure at which the
piston floats.
With a PG7302 at low pressure there may
be excessive air in the test system.
Purge the test system of air prior to operating
the PG7302. Use the PPCH prime and
purge function. PPCH Operation and
Maintenance Manual
Automated pressure control (GEN) is
extremely slow or never floats the
pistons.
The test volume is too large or the volume
setting is incorrect (PG7302/PG7307 only).
Reduce the test volume and/or, if possible,
make the tolerance setting smaller, adjust
the volume setting (PG7302/PG7307 only).
3.9.9.4, 3.9.9.6
Automated pressure control (GEN) is
not increasing or decreasing pressure
as expected.
Pressure or vacuum supply on the
pressure controller is not adequate.
Provide correct pressure and vacuum
supplies to pressure controller. See the
pressure controllers Operation and
Maintenance Manual.
Automated pressure control (GEN) is
not starting after setting a target and
loading mass AND measurement
mode is absolute by vacuum.
In absolute by vacuum measurement
mode, fefore automated pressure
generation begins, the reference vacuum
must be below 20 Pa.
Wait for vacuum to reach less than 20 Pa,
check and correct vacuum supply if
inadequate.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...