2. INSTALLATION
Page 31
© 1998-2008 DH Instruments, a Fluke Company
2.4.8
CHECK PROPER OPERATION OF PISTON BEHAVIOR
MEASUREMENTS
Float the piston and rotate it (see Sections 2.4.6 and 2.4.7).
Press
[SYSTEM]
once to reach the first SYSTEM run screen. Verify that the piston position,
piston fall/rise rate, piston rotation rate and piston rotation decay rate are indicating correctly
(see Section 3.9.5). The piston position reading system may be calibrated using an on-board
procedure if necessary (see Section 5.2.2). Calibration of piston position is recommended
when installing a new PG7000 and regularly after installation.
Press
[SYSTEM]
again to reach the second SYSTEM run screen. Verify that the piston-cylinder
temperature and temperature rate of change are indicating correctly (see Section 3.9.5).
If checking a PG7601 Platform, verify that the vacuum reference measurement is operating
correctly (see Section 2.4.8.1).
2.4.8.1
VERIFY VACUUM REFERENCE (PG7601 ONLY)
PG7601 includes provisions for establishing and measuring a vacuum reference.
To verify the vacuum reference capability:
n
Install a piston-cylinder module (see Section 2.3.2).
o
Install the bell jar on the PG7601 Platform (the bell jar aligns on the PG7601
vacuum plate and seals itself). If using AMH-38 automated mass handling,
install the AMH-38 mass handler (see the AMH-38/AMH-100 Operation and
Maintenance Manual).
p
Apply a vacuum through the reference vacuum port (KF25 port on front left
side of platform or KF40 if using the optional AMH-38 automated mass
handler).
q
Press
[SYSTEM]
twice to observe the value of vacuum read by the
PG7601’s built-in vacuum gauge.
If a vacuum pump of adequate capacity has been correctly connected to the
reference vacuum port, the vacuum read by the PG7601 built-in vacuum gauge
should go to 4 Pascal [Pa] or lower in less than five minutes on the first pump
down and two to three minutes on immediately subsequent pump downs.
If this performance is NOT achieved:
•
The vacuum pump may be inadequate.
•
The connection of the vacuum pump to the PG7601 platform of AMH-38 may
have leaks or excessive restrictions.
•
The PG7601 built-in vacuum gauge or external vacuum gauge may be
incorrect.
•
There may be a leak in the PG7601.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...