5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION
Page 175
© 1998-2008 DH Instruments, a Fluke Company
5.2.1.2
BAROMETRIC PRESSURE SENSOR
PURPOSE
To view and adjust the output of the on-board barometric sensor.
OPERATION
To view the output of the barometric pressure sensor, press
[SPECIAL]
and
select
<1atmp>
,
<1view>
. The display is:
1. Current reading of the barometric sensor in
current pressure unit of measure.
2. Indication that this is a barometric pressure
sensor display.
3. Current value of the Pressure Multiplier
applied to the barometric pressure sensor
reading.
4.
Current value of the Pressure Adder (always
in Pascal [Pa]) applied to the barometric
pressure sensor reading.
100.0031 kPa atmP
PA 101 Pa PM0.99999
To adjust the values of PA and/or PM press
[SPECIAL]
and select
<1atmP>
,
<2cal>
to access a screen in which the values of PA and PM can be edited.
From here, press
[ESCAPE]
and select
<1atmP>
,
<1view>
to view the
barometric sensor reading with the edited calibration coefficients applied.
See Section 5.2.1.1 for an explanation of Adders and Multipliers and their use
in adjusting internal sensors.
5.2.1.3
AMBIENT TEMPERATURE SENSOR
PURPOSE
To view and adjust the output of the on-board ambient temperature sensor.
The ambient temperature sensor and piston-cylinder module temperature
sensor are of the same technology and read in the same manner. However, their
recommended calibration procedures differ reflecting the different measurement
uncertainty specifications of the two measurements. See Section 5.2.1.5 for
recommended calibration procedure on the piston-cylinder module temperature
sensor, which is used to measure temperature with much lower uncertainty.
OPERATION
To view the output of the ambient temperature sensor, press
[SPECIAL]
and
select
<2ambT>
,
<1view>
. The display is:
1.
Current reading of the ambient temperature sensor.
2. Indication that this is an ambient temperature
display.
3.
Current value of the Temperature Multiplier applied
to the ambient temperature sensor reading.
4.
Current value of the Temperature Adder (always in
°C) applied to the ambient temperature sensor
reading.
20.51 °C ambT
TA 0.1 °C PM 0.99997
To adjust the values of TA and/or TM press
[SPECIAL]
and select
<2ambT>
,
<2cal>
to access a screen in which the values of TA and TM can be edited.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...