PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 106
Some screens, such as the SPECIAL menu, go beyond the two lines provided by
the display. This is indicated by a flashing arrow in the second line of the display.
Press the [
←
] and [
→
] keys to move the cursor to access the lines that are NOT
visible or directly enter the number of the hidden menu choice if you know it.
Refer to Sections 3.11.1.6, PRINCIPLE and 3.6 for information on mass
group definitions. To create the mass set, each and every mass in the mass set,
no more and no less, (but not the piston and the bell) must be identified and its
true mass entered. To select a mass group press the
[
←
]
and
[
→
]
keys to
position the cursor on the mass group to be edited. If the mass is needed but not
shown, select
<ADD>
to create a new mass group. To delete a mass group,
enter
<0>
as the number of masses in the group.
All standard PG7000 MANUAL mass sets have a makeup mass (MKUP). This
mass may be 4, 4.5 or 9 kg depending on the mass set (see Section 3.6, PRINCIPLE).
It is imperative that the make up mass be defined and entered in the MKUP group.
Operation within each mass group is identical, for example, select
<5.00>
. The
display is:
1.
Edit field for number of masses in the mass
group.
# of masses in group
(0 deletes grp): 1
Edit the number of masses in group to reflect the number of masses of that
nominal value there are in the group and press
[ENTER]
. The display is:
1.
Sequential ID number of this specific mass
within the mass group.
2. Edit field for the true value of this specific
mass.
5.0 kg mass #1
True mass: 5.00000
Edit the mass value to the true value of the specific mass identified and press
[ENTER]
.
For high line differential mode, when setting up the tare mass set of the tare
PG7000, the true mass of each mass is entered as equal to the nominal value. Enter
the nominal mass values for the true mass values.
If there are additional masses in this mass group, the next display is the same as
the display immediately above but with the next mass specific mass sequential
ID number in the mass group. The screens continue until the true value of all of
the masses within the group has been entered. After the last entry, the screen
returns to the
<Select mass grp>
screen.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...