3. GENERAL OPERATION
Page 95
© 1998-2008 DH Instruments, a Fluke Company
To edit an existing SETUP file or create a new SETUP file, press
[SETUP]
and
select
<3edit>
. The display is:
1.
Entry field for number of SETUP file to be edited or created.
Defaults to SETUP file currently selected. Must be a number
from 2 to 10.
Edit SETUP file:
#02
Enter the desired SETUP file number and press
[ENTER]
.
If a valid SETUP file number is entered, the display is:
Edit #02 latmP 2ambT
3%RH 4PCT 5g 6READY
7vac
Selecting an item leads to the variable source choices menu for that item. The cursor is on the
source choice that is currently selected in that SETUP file number. Select the desired
variable source. The selection causes the variable choice selection to be made and returns
to the edit menu. From the variable source choice menus, pressing
[ENTER]
or
[ESCAPE]
returns to the edit selection menu. This allows easy, discreet movement between variable
source choices when editing a SETUP file.
Select
<1atmP>
to specify the atmospheric pressure variable source for the pressure values
that are used to calculate ambient air density and to add to gauge pressure measurements to
calculate absolute pressure in absolute by addition of atmosphere mode. Selecting
<1atmP>
offers the atmospheric pressure variable source choices:
<1internal>
Use real time readings from PG7000’s on-board barometer.
<2normal>
Use a fixed value of 101.325 kPa (14.6959) psi.
<3user>
Use a fixed user entered value. If
<3user>
is selected, the user value must be
entered, in the current pressure unit of measure (see Section 3.9.3).
<4COM2>
Use real time readings from the device connected by RS232 to PG7000’s COM2
port. The external barometer must be properly set up to communicate with PG7000
(see Section 3.11.5.4 for information on setting up to read an external barometer).
For differential mode operation (see Section 3.9.4) the AtmP selection must be <4COM2>.
Select
<2ambT>
to specify the ambient temperature variable source for the temperature
values that are used to calculate ambient air density. Selecting
<2ambT>
offers the
temperature variable source choices:
<1internal>:
Use real time readings from PG7000’s on-board ambient temperature platinum
resistance thermometer (on the platform rear panel).
<2normal>:
Use a fixed value of 20
°
C.
<3user>:
Use a fixed user entered value. If
<3user>
is selected, the user value must be
entered, in degrees Centigrade
[
°
C]
.
Select
<3%RH>
to specify the relative humidity variable source for the relative humidity
values that are used to calculate ambient air density. Selecting
<3%RH>
offers the relative
humidity variable source choices:
<1internal>:
Use real time readings from PG7000’s on-board relative humidity sensor (on the
platform rear panel).
<2normal>:
Use a fixed value of 45 %RH.
<3user>:
Use a fixed user entered value. If
<3user>
is selected, the user value must be
entered, in percent relative humidity
[%RH]
.
Select
<4PCT>
to specify the piston-cylinder temperature variable source for the temperature
values that are used to compensate piston-cylinder effective area for temperature. Selecting
<4PCT>
offers the temperature variable source choices:
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...