PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 124
The external vacuum gauge parameters including the communication string can
also be set by remote command which is more convenient than front panel entry (see
Section 4.3.4.2, “UDV” command).
After the string value has been entered
the display is:
Reply conv coef:
1.000000 Pa/unit
<conv coef>
is the conversion coefficient that PG7600 will use to convert
readings from the external vacuum gauge to the pressure unit of measure
Pascal
[Pa]
. If the readings from the external vacuum gauge are NOT Pascal,
edit the conversion coefficient value as needed to convert the readings to Pascal.
When the conversion coefficient is entered
the next setup screen is opened. The
display is:
Leading characters to
Strip: 0
This entry specifies the leading characters in the return string that precede the
numerical value of the pressure. PG7601 will strip the specified number of
characters from the front of the return string and assume that the next character is
the first number defining the current value of reference vacuum.
To verify whether external vacuum gauge communications have been set up
properly, select a SETUP file that specifies COM2 under Vac (see Section 3.10) and
verify that communications with the external vacuum gauge are achieved. Then
view current vacuum readings in the AMBIENT run screen (see Section 3.9.6) and
verify that the readings are correct.
For communications with an external vacuum gauge on COM2 to be established
properly, the inquiry string to send to the vacuum gauge must be defined following
this section and the PG7000 COM2 port and vacuum gauge COM port must have the
same communications settings (or if the vacuum gauge is connected to COM2 of a
DHI RPM, the RPM3 COM2 and vacuum gauge COM port must have the same
communication settings. Press [SPECIAL] and select <5remote>, <2COM2>,
<1settings> to set up PG7000 COM2 communications settings. When unable to
establish communications between COM2 and the external vacuum gauge, consider
using the PG7000 RS232 self test to verify COM2 operation (see Section 3.11.5.3).
3.11.6
<6gl>
PURPOSE
To view and set the value of local gravity used by PG7000 in reference pressure calculations
(see Section 7.2) when
gl
is specified as the gravity value in SETTINGS (see Section 3.10).
OPERATION
To access the PG7000 local gravity value (gl), press
[SPECIAL]
and select
<6gl>
. The display is:
Local gravity:
9.806650 m/s2
The value displayed is the PG7000 local gravity value. This value can be edited unless it is
protected by the current security level. The default value is normal gravity of 9.806650 m/s
2
.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...