3. GENERAL OPERATION
Page 65
© 1998-2008 DH Instruments, a Fluke Company
After high line differential mode is initialized,
the display
of the tare PG7000
is:
1.
Standard Ready/Not Ready indicators for the
tare PG7000.
2.
Last line pressure set in the pressure unit of
measure that was active when the line
pressure was set. The unit of measure
remains the same, even if the reference
PG7000’s unit of measure is changed, until a
new line pressure is set.
< 1000.70 psi LP
- 3.7 30.3 kg + trim
3.
Label indicating that the mass load includes whatever trim masses were loaded on the tare
piston in the crossfloating process to reach equilibrium.
4.
Tare mass load for the last line pressure set.
5.
Current piston position of the tare PG7000.
While in high line differential pressure mode, the
tare
PG Terminal keypad
is inactive.
After high line differential mode is initialized,
the display
of the reference PG7000
is:
1.
Current (last set) value of line pressure in
current pressure unit of measure.
Line P: 1000 psi
New line P? 1no 2yes
Select
<1no>
to use the existing line pressure settings and return to the previous
run screen in high line differential measurement mode (see Section 3.9.4.2,
Setting Differential Pressures at a High Line Pressure
).
Select
<2yes>
to set a new line pressure value or repeat the line pressure
setting procedure at the current line pressure value (see Section 3.9.4.2,
Setting
a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)
).
The line pressure setting procedure must be executed each time the line
pressure is changed and should be executed at the beginning of each differential
pressure calibration sequence even if the line pressure is the same as the line
pressure previously used. Select <2yes> to go to the line setting procedure.
Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)
To meet the full performance potential of PG7000 high line differential mode
operation, external influences on the piston gauges must be reduced or eliminated.
Air currents and vibrations are the most significant possible influences. Do not
operate near an active air conditioning or heating duct, avoid opening and closing
doors or any movement of personnel around the system.
Leaks at any point in the test system are highly detrimental to measurement
results in high line differential mode. Thoroughly leak check the system before
operation and correct any leaks detected.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...