PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 176
From here, press
[ESCAPE]
and select
<2ambT>
,
<1view>
to view the ambient
temperature sensor reading with the edited calibration coefficients applied.
See Section 5.2.1.1 for an explanation of Adders and Multipliers and their use
in adjusting internal sensors.
5.2.1.4
RELATIVE HUMIDITY SENSOR
PURPOSE
To view and adjust the output of the on-board relative humidity sensor.
OPERATION
To view the output of the relative humidity sensor, press
[SPECIAL]
and select
<3%RH>
,
<1view>
. The display is:
1. Current reading of the relative humidity
sensor.
2. Indication that this is a relative humidity
display.
3.
Current value of the Humidity Multiplier applied
to the relative humidity sensor reading.
43.2 %RH amb%RH
HA 5 %RH HM 0.99984
4.
Current value of the Humidity Adder (always in %RH) applied to the relative humidity sensor
reading.
To adjust the values of HA and/or HM press
[SPECIAL]
and select
<3%RH>
,
<2cal>
to access a screen in which the values of HA and HM can be edited.
From here, press
[ESCAPE]
and select
<3%RH>
,
<1view>
to view the relative
humidity sensor reading with the edited calibration coefficients applied.
See Section 5.2.1.1 for an explanation of Adders and Multipliers and their use
in adjusting internal sensors.
5.2.1.5
PISTON-CYLINDER MODULE TEMPERATURE SENSOR
PURPOSE
To view and adjust the output of the on-board piston-cylinder module
temperature sensor.
PRINCIPLE
Piston-cylinder module temperature is measured by a platinum resistance
thermometer (PRT) built into the PG7000 mounting post. The resistance of the
PRT is read by PG7000's internal ohmic measurement system and temperature
is calculated as follows:
Temperature = (Measured resistance - resistance at 0
°
C)/slope
The PRT is a 10052 4-pole type following DIN Norm 43760. The current
supplied to the PRT is 1 mA and the slope of resistance relative to temperature
over the valid temperature range of 0 to 40
°
C is 0.3896
Ω
/
°
C. The uncertainty
specification is ± 0.1
°
C (
±
0.02
Ω
) with a resolution of 0.01
°
C.
The ohmic measurement system in PG7000 is self-calibrated using on-board
100 and 110
Ω
reference resistors.
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...