PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 82
OPERATION
To adjust the AutoGen piston float target press
[GEN]
,
<2target>
. Edit the value of
the target as desired. The default is + 1 mm (0 mm for PG7202). The entry should
not 2.5 mm.
3.9.9.2
<3RAISE>
PURPOSE
To turn ON and OFF a function that causes the automated generation component (see
Section 3.9.9.) to raise the PG7000 piston to the top of its stroke before automated
mass manipulation by an AMH automated mass handler. This can avoid a large
pressure change which may occur in certain circumstances when AMH lifts the mass
load off of the piston.
OPERATION
To turn ON and OFF the raise piston before mass loading fuction, press
[GEN]
,
<3raise>
. The cursor is on the choice corresponding to the current state. Select
<1no>
for the piston not to be raised or
<2yes>
to raise the piston. The default is
<1no>
.
3.9.9.3
<4UL>
PURPOSE
To read and/or set the UPPER LIMIT (UL) of the automated pressure generation
component used by AutoGen. This function is used to protect against accidential
overpressure when using the PG7000 automated pressure generation function (see
Section 3.9.9 and the generation component’s Operation and Maintenance Manual,
UL section)
OPERATION
To view or set the UPPER LIMIT of the automated pressure control component
used by AutoGen, press
[GEN]
,
<4UL>
.
If AutoGen is not currently ON, the control component UPPER LIMIT cannot be
accessed and an error message is displayed.
If AutoGen is ON, the current UPPER LIMIT of the automated control component
is displayed and can be edited.
The automated pressure control component will abort pressure generation and
beep repeatedly if its UPPER LIMIT is exceeded.
3.9.9.4
<5TOL>
PURPOSE
To read and/or set the pressure measuring tolerance of the pressure controller used by
AutoGen to automate pressure control.
This function is used only when the pressure controller is a PPC3.
PRINCIPLE
The control tolerance function can be used to reduce the time required to set pressure
and float the PG7000 piston when the pressure controller used is significantly more
accurate than the default tolerance value. The tolerance can also be described as the
degree of agreement between the pressure control’s pressure measurement and the
pressure value set by the PG7000 when its piston is floating. In general, making the
tolerance smaller reduces the time required to float the piston and making the
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...