PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 66
Setting a line pressure has two steps:
Setting the line pressure and determining the natural fall rates of the tare and
reference pistons.
Performing a crossfloat of the pistons to reach equilibrium and minimize
differential pressure zero offset.
To set a line pressure press
[MODE], <3HLdif>, <1run>, <2yes>
on the
reference
PG7000’s PG Terminal. The display is:
1.
Edit field for entry of desired line pressure
value in current pressure unit of measure.
Defaults to last value used.
Target line P:
1000 psi
Values of line pressure are always entered and displayed in gauge mode
(i.e., relative to atmospheric pressure). Line pressure mass loading instructions are
always given with 0.1 kg resolution. The minimum line pressure is the pressure resulting
when the mass of the mass bell is loaded.
Enter the desired line pressure value.
The next display indicates the mass to
load to set the line pressure.
Load nominal mass on
T & R: 33.3 kg
This display corresponds to the standard mass loading instruction display of
pressure to mass
mode (see Section 3.9.11.1). The mass loading instruction is
always given with 0.1 kg resolution regardless of the current mass loading
resolution setting (see Section 3.9.10). Load the indicated mass on both the
tare
and
reference
PG7000s following PG7000 mass loading protocol using main
and fractional masses only (do not use trim masses) (see Section 3.6). Press
[ENTER]
when ready.
Line pressure mass instructions are always given with 0.1 kg resolution. This is
to preserve the smaller masses for setting the differential pressure. The differential
pressure mass loading instructions follow conventional PG7000 mass loading
protocol using the masses that remain after the line pressure has been set. It may
be necessary to select the line pressure value to assure that adequate masses
remain to set the desired differential pressure. For example, if the line pressure
requires loading 33.3 kg, a 0.2 and 0.1 kg mass will be used to set the line pressure
and will not be available if needed to set the desired differential pressure. If this
situation occurs, consider changing the line pressure so that it is set using a whole
number of kilograms, in this example 33 or 34 kg.
The display is:
Open bypass, float T
Close bypass,[ENTER]
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...