4. REMOTE OPERATION
Page 161
© 1998-2008 DH Instruments, a Fluke Company
SETUP(=)
Purpose
To change or read the current SETUP to be used.
Syntax “SETUP=n”
“SETUP”
Default “SETUP=1”
Argument N/A
Remarks
Up to 19 SETUP files can be created by the user to allow quick change from one
configuration of pressure calculation sources to another. Each configuration can specify
the source of the ambient measurements, the gravity, and the
Ready/Not Ready
criteria.
The “AMBHx”, “AMBPx”, “AMBTx”, “READYx”, “VACP” and “PCTX” commands
reference the SETUP command. The first setup (“SETUP=1”) is reserved for the factory
default setup, and cannot be changed. Setup #21 is reserved for remote access only.
Selecting a setup that references an external device (barometer or vacuum gauge) can
take up to 5 seconds to reply.
Example Typical
command:
Typical reply:
“SETUP=2”
“2”
Error
See 9.3.2 Error Messages
ERR #1
If setup is <1 or >21
See Also
3.10, AMBHx”, “AMBPx”, “AMBTx”, “READYx”, “VACP”, “PCT
SN
Purpose
To read the PG7000 serial number.
Syntax “SN”
Default N/A
Argument N/A
Remarks
Each PG7000 is assigned a serial number. This command reads this number back.
Example Typical
command:
Typical reply:
“SN”
“201”
Error None
SPEED
Purpose
To read the current piston rotation speed and rotation decay (deceleration).
Syntax “SPEED”
Default N/A
Argument N/A
Remarks
The current piston rotation speed and decay rate are returned separated by a comma.
The decay rate is returned as ‘0’ if there is not enough valid data to calculate the rate.
The speed will also be returned as ‘0’ if the bell is not installed.
Example Typical
command:
Typical reply:
“SPEED”
“44.2 rpm, 2.3 rpm/min
Error None
See Also
3.9.5
TIME(=)
Purpose
To read or set the internal real time clock.
Syntax “TIME=HH:MMxp”
“TIME”
Default N/A
Argument
The time is in the HH:MMxp format.
HH:
MM:
xp:
The hour 1 to 12
The minute 00 to 59
“am” or “pm”
Remarks
The internal time clock is part of the internal calendar. The seconds are always set to
“00” when setting the time.
Example Typical
command:
Typical reply:
“TIME=8:31PM”
“8:31pm”
Error
See 9.3.2 Error Messages
ERR #1
If invalid time argument
See Also
3.11.4.3
UCOEF
Purpose
To read the conversion coefficient used to convert Pascal to the current pressure unit of
measure.
Syntax “UCOEF”
Default N/A
Argument N/A
Remarks
Use this command to read the pressure conversion coefficient used to convert pressure
in Pascal to the displayed pressure units. Pressure in Pascal multiplied by this
coefficient yields the pressure in the PG7000 units.
Example Typical
command:
Typical reply:
“UCOEF” (current units are “kPa”)
“1.000000e-003”
Error None
See Also
7.1.1, 3.9.3, UNIT
Содержание PG7000
Страница 10: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 44: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 34 N NO OT TE ES S ...
Страница 182: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 172 N NO OT TE ES S ...
Страница 210: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 200 N NO OT TE ES S ...
Страница 214: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 204 N NO OT TE ES S ...
Страница 224: ...PG7000 OPERATION AND MAINTENANCE MANUAL 1998 2008 DH Instruments a Fluke Company Page 214 N NO OT TE ES S ...